AGFA NDT FIXCLEAN is a one-piece care product that provides cleaning of fixation zones, intermediate rinsing and final rinsing in developing machines. It can be used to clean fixation tanks in mixers and manual developing systems.
Instructions for use:
REMEMBER: FIXCLEAN should never be mixed with fixer waste.
1. Empty and thoroughly flush the tank, rack and recirculation system with clean water to remove all traces of fixer.
2. Do not remove racks.
3. Fill each tank with water to approximately 80% capacity.
4. Slowly add the FIXCLEAN concentrate, avoiding splashing, and turn on the recirculation pump (for mixing).
With a tank for fixing and / or water less than 10 liters:
Pour about 5 liters of water into the tank (about 35 ° C).
Add 1 liter of FIXCLEAN concentrate to the tank.
Top up with water in the tank until it overflows.
5. Empty the bottle from under FIXCLEAN should be rinsed with water, after which this can be added to the working solution in the tank.
6. Add water to the reservoir until it is filled by turning on the recirculation pump.
7. Cleaning time when the recirculation pump is running:
45 min at temperature ≤ 30 ° C
30 min at a temperature> 30 ° C
If the amount of sludge in the washing zone is very large, you can increase the concentration of FIXCLEAN. If necessary, you can apply FIXCLEAN in a concentrated form.
Always follow the instructions and follow the safety instructions, especially when working with 100% concentrate.
8. When cleaning other containers for fixation, leave FIXCLEAN for at least 15 minutes. In these cases, you can use large concentrations, depending on the degree of contamination.
9. Empty each tank completely. Drain the FIXCLEAN waste solution to the developer waste.
10. Before filling the tank with fresh tap water (about 15 ° C), rinse the tank and stand with running tap water. Drain the wash water to the developer waste.
11. Fill the tank with fresh tap water (about 15 ° C) and turn on the recirculation pump for 15 minutes.
12. Empty each tank to the fixture drain. Repeat the cleaning and rinsing, if necessary, until the tanks and racks are clean.
13. Cleaning completed. Close the drain hole of each of the tanks, then pour fresh fixer solution into the fixation tank and tap water into the water tank.
Fixclean Development Cleaners